| Model Number: | 3HYE |
| Place of Origin: | UK |
| MOQ: | 1 |
| Payment Terms: | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
| Delivery Time: | 3-5workingdays |
| Nominal Range: | 0-10 000ppm |
| Maximum Overload: | 20 000ppm |
| Expected Operating Life: | Two years in air |
| Output Signal: | 0.003 ± 0.001 µA/ppm |
| Resolution: | 10ppm |
| Temperature Range: | -20°C to +50°C |
| Pressure Range: | Atmospheric ± 10% |
| Pressure Coefficient: | 0.006 % signal/mBar |
| T90 Response Time: | <70 seconds |
| Relative Humidity Range: | 15 to 90% non-condensing |
| Typical Baseline Range (pure air): | +2 to -150ppm equivalent |
| Company Info. |
| ShenzhenYijiajie Electronic Co., Ltd. |
| Verified Supplier |
| View Contact Details |
| Product List |
3HYE Hydrogen 3 series electrochemical gas sensor 0-10000ppm
The output of a CiTiceL can vary with
temperature. The graph here shows the variation
in output with temperature for 3HYE CiTiceLs
based on a sample of about 16 sensors. The
results are shown in the graph as a mean for the
batch, and expressed as a percentage of the
signal at 20°C.
From a statistical viewpoint, for a sample of
this size, the range in values observed for all
sensors of this type will fall within a range three
times the standard deviation above or below the
mean. Assuming therefore this sample is typical,
then the temperature behaviour of all 3HYE
CiTiceLs will fall in the band +3SD to -3SD.
| Maximum Zero Shift (+20°C to +40°C) | -150ppm equivalent |
| Long Term Output Drift | <2% signal loss/month |
| Bias Voltage | Not required |
| Repeatability | 2% of signal |
| Weight | 22g |

|