Power: 8.4V lithium battery (built-in)
Interference voltage: <20V (should be avoided)
Interference current: <2A (should be avoided)
Test frequency: 128Hz
Short circuit test current: > 20mA (Sine Wave)
Electrode spacing range: AC 28V max
  • Haven't found right suppliers
  • Our buyer assistants can help you find the most suitable, 100% reliable suppliers from China.
  • And this service is free of charge.
  • we have buyer assistants who speak English, French, Spanish......and we are ready to help you anytime!
Submit Buying Request
  • Product Details
  • Company Profile

ES3001 Soil Resistivity Tester for Accurate Measurements

Product Information


Soil resistivity tester also known as grounding resistance-soil resistivity tester precision grounding
resistance tester, etc. is a common instrument for inspection and measurement of grounding
resistance commonly used instruments, the use of a large LCD grey and white screen backlit
display and microprocessor technology to meet the two, three, four-wire test resistance and Shi soil
resistivity requirements. Applicable to telecommunications, electric power, meteorology, engine
room, oil field power distribution lines, tower transmission lines, petrol stations, factories, grounding
network lightning rods and so on. Instrument testing precision, fast, simple, stable and reliable
characteristics.


Technical Parameter

FunctionTwo three four-wire measure grounding resistance, soil resistivity;Ground voltage, AC voltage measurement
Ambient temperature and humidity23℃±5℃,below 75%rh
Power8.4V lithium battery (built-in)
Interference voltage<20V (should be avoided)
Interference current<2A (should be avoided)
Measure R electrode spacinga>5d
Measured electrode spacinga>20h
Auxiliary ground resistanceReference condition <100Ω, working condition <5kΩ
RangeGrounding resistance: 0.00Ω~30.00kΩ Soil resistivity: 0.00Ωm~9999kΩm Grounding voltage: 0.00V~100.0V
measurement modePrecise four-wire, three-wire measurement, simple two-wire measurement
Measurement methods

Grounding resistance: rated current change pole method

Soil resistivity: four-pole method

Ground Voltage: Average Rectification(between S-ES interface)

Test frequency128Hz
Short circuit test current> 20mA (Sine Wave)
Open circuit test voltageAC 28V max
Electrode spacing rangeCan be set 1m ~ 100m
Change gearfully automatic shifting 0.00Ω ~ 30.00kΩ;fully automatic shifting 0.00Ωm~9000kΩm
BacklightControllable gray screen backlight, suitable for use in dim places
Display mode4-bit large LCD display, gray screen backlight
Measurement instructionsLED flashing indicator during measurement
LCD size111mm×68mm
LCD display field108mm×65mm
Instrument size277.2mm×227.5mm×153mm
Standard test line length4 strips:red 15m,black 15m, yellow 10m,green 10m each one
Simple test line2strips:yellow 1.6m,green 1.6m each one
Auxiliary Grounding rod4PCS:φ10mm×200mm
Measure timeGround voltage:about 3 times/sec grounding resistance,soil resistivity: about 7 seconds/time
Line voltageMeasurement below AC100V (ground voltage measurement function cannot be used to measure commercial power)
USB interfaceWith USB interface, software monitoring, storage data can be uploaded to the computer, save and print
Communication LineOne USB communication line, 1.5m long
Data storage500 groups, "MEM" storage indicates, flashing "FULL" symbol indicates that the memory is full
Data reviewData review function: "MR" symbol display
Overflow displayOver-range overflow function: "OL" symbol display
Interference testAutomatic identification of interference signals, "NOISE" symbol indication when the interference voltage is higher than 5V
Auxiliary grounding testWith auxiliary ground resistance test function, 0.00KΩ ~ 30kΩ (Rh max = 3kΩ +100R <50kΩ; Rs max = 3kΩ +100R <50kΩ)
Alarm functionAlarm when the measured value exceeds the alarm setting value
Battery voltageReal-time display of battery power, reminding timely charging when battery voltage is low
Automatic Shutdown“APO” Indicates, automatic Shutdown After 15 Minutes
Power consumptionStandby: 40mA Max(Backlight off) Turn on backlight: about 43mA measuring:75mA Max(Backlight off)
WeightInstrument: 2450g(including battery) Test lines:1300g Auxiliary grounding rod: 850g(4PCS)
Working temperature and humidity-10℃~40℃;below 80%rh
Storage temperature and humidity-20℃~60℃;below 70%rh
Overload protectionGrounding resistance:AC 280V/3 seconds between H-E and S-ES ports
Insulation resistance>20MΩ(500V between circuit and housing)
Pressure resistanceAC 3700V/rms(between circuit and housing)
Electromagnetic propertiesIEC61326(EMC)
Suitable for safety regulations

IEC61010-1(CAT Ⅲ 300V,CAT IV 150V,pollution level 2);

IEC61010-031;

IEC61557-1(grounding resistance);

IEC61557-5(soil resistivity);

JJG 366-2004.