dead weight tester pressure gauge calibration
Description dead weight tester pressure gauge calibration
The Piston Dead Weight Tester, used for pressure calibration, is a
primary standard apparatus employing a piston-cylinder mechanism.
It establishes equilibrium between the applied load and the
pressure beneath the piston. This device is considered a primary
standard as its pressure readings are based on fundamental
quantities like length, mass, and time. It is primarily used in
calibration labs for calibrating pressure transfer standards,
including electronic pressure measuring devices. By providing a
reference pressure traceable to basic physical quantities, it
ensures the accuracy and reliability of pressure measurements
across various applications.
Detailed Images dead weight tester pressure gauge calibration

Specification dead weight tester pressure gauge calibration
Item | | YS6 | YS60 | YS250 | YS600 |
Range | MPa | 0.04-0.6 | 0.1-6 | 0.5-25 | 1-60 |
Up limit of measurement | MPa | 0.6 | 6 | 25 | 60 |
Down limit of measurement | MPa | 0.04 | 0.1 | 0.5 | 1 |
Piston nominal area | cm² | 1 | 0.5 | 0.1 | 0.05 |
Chassis and piston | Nominal mass kg | 0.4 | 0.5 | 0.5 | 0.5 |
Generatedpressure MPa | 0.04 | 0.1 | 0.5 | 1 |
Special weight | Nominal mass kg | 0.1;0.5 | 0.5;2.5 | 0.5;2.5 | 0.5;2.5 |
Generatedpressure MPa | 0.01;0.05 | 0.1;0.5 | 0.5;2.5 | 1;5 |
Number | 6;10 | 4;11 | 4;9 | 4;11 |
Both ends of the lock nut | | M20×1.5 | M20×1.5 | M20×1.5 | M20×1.5 |
Weight | kg | 35 | 60 | 55 | 90 |
Working liquid | 25#Transformer oil and aviation kerosene mixed (4.39:1).20 ℃
Kinematic viscosity 9 ~ 12 centistokes, The acid value of less than 0.05 mg KOH/ G | sebacate 20 ℃ kinematic viscosity 20-25 centistokes, acid value is
not greater than 0.05 mg KOH/ G |
|
Advantages dead weight tester pressure gauge calibration
Feature china dead weight tester pressure gauge calibration
1. The piston and cylinder are made of high strength, high
stiffness, and low temperature coefficient of linear expansion of
the carbide grinding tungsten carbide material , with a wide range
of ambient temperature , stable performance 20 ± 5 ℃.
2. The load center of gravity low, weight by Cradle applied
directly to the piston, the piston smooth operation, the pressure
fluctuation is extremely small, accurate indication.
3. When pressure more than 25MPa ,the medium uses a small viscosity
isooctyl sebacate, thus ensuring a very high sensitivity.
4. The weights used are all non-magnetic stainless steel material.
5.The piston displacement sensor used to monitor the working
position, the piston up and down stroke of ±1.5mm, instrument
display is sensitive, smart, and accurate.
6. Wide range, at the lower 1MPa, the basic error of ± 0.02% or ±
0.05%, a KY100 type manometer, capable of reaching the upper limit
of the current domestic 6,60,100MPa three measuring manometer.
7. pressure regulator easy , even at 250MPa pressure could ease.